Osciloscópio Portátil AEMC OX 9304
Modelo: OX 9304
Large touch screen 7” color LCD display
Ideal for both laboratory and field use, the Model OX 9304 utilizes the patented system of “plug-and-play” accessories, individual isolation of each of the measurement channels, and a range of remote management capabilities. The Ethernet link and Wi-Fi lets you remotely control the oscilloscope without loading any software to your PC. Android App available to download from Google play store.
TECHNICAL SPECIFICATIONS
Type of display: 7" WVGA color TFT LCD touch screen, 800x480 – LED backlighting (adjustable standby mode)
Different display mode: 2,500 real acquisition points on screen - Vectors with interpolation
Display of curves on screen: 4 curves + 4 references – Split Screen & Full Screen modes
Screen commands: Touch screen – ANDROID-type icons and graphical commands – customizable channel colors
Choice of language: 15 complete languages, menus & online help
OSCILLOSCOPE MODE
Vertical deflection
Bandwidth: 60 MHz (OX 9062); 100 MHz (OX 9102, OX 9104); 300 MHz (OX 9304); 15 MHz, 1.5 MHz or 5 kHz bandwidth limiter
Number of channels: 2 isolated channels (OX 9062, OX 9102); 4 isolated channels (OX 9104, OX 9304);
Input impedance: 1 MΩ ± 0.5% , approx. 12 pF
Maximum input voltage: 600 V / CAT III (1,000V per Probix) – from 50 to 400 Hz – Probix safety connectors
Vertical sensitivity: 16 ranges from 2.5 mV to 200 V/div and up to 156 μV/div in vertical zoom mode (12-bit converter) – Accuracy ± 2%
Vertical zoom: "One Click Winzoom" mode (12-bit converter and direct graphical zoom on screen) – x 16 max.
Probe factor (non-Probix): 1 / 10 / 100 / 1,000 or any scaling – definition of measurement unit
Horizontal deflection
Sweep speed: 35 ranges from 1 ns/div to 200 s/div., accuracy ± [50 ppm + 500 ps] – Roll mode from 100 ms to 200 s/div
Horizontal zoom: "One Click Winzoom" system (direct graphical zoom on screen) x 1 to x 5 or x 100 – storage 100 kpts/channel
Triggering
Mode: On all the channels - automatic, triggered, one-shot, auto level 50%
Type: Edge, pulse width (16 ns-20 s), delay (48 ns to 20 s), counting (3 to 16,384 events) Continuous adjustment of Trigger position
Coupling: AC, DC GND, HFR, LFR, noise – Level and Hold-Off adjustable from 64 ns to 15 s
Sensitivity: ≤ 1.2 division p-p up to 300 MHz
Digital storage
Maximum sampling rate: 2.5 GS/s in one-shot mode on each channel (100 GS/s max. in ETS mode)
Vertical resolution: 12 bits (vertical resolution 0.025 %)
Memory depth: 100 kpts per channel and file viewer in the manager
User storage, File management: Internal = 1 GB to store the files: trace, text, configuration, math functions, System memory: .pdf print files, .png image files+ high-capacity removable μSD-Card: SD 2 GB, SDHC 4-32 GB and SDXC > 32 GB
GLITCH mode: Duration ≤ 2 ns – 500,000 Min/Max pairs
Display modes: Envelope, vector, accumulation-, averaging (factors 2 to 64) – XY (vector) and Y(f)=FFT
Other functions
AUTOSET: Complete in under 5 s, with recognition of the channels – Frequency > 30 Hz
FFT analyzer & MATH functions: 2,500-point FFT (Lin or Log) with measurement cursors – Functions + , - , x , / and mathematical function editor
Cursors: 2 or 3 cursors: simultaneous V and T with AUTO measurement: T1, T2, Dt, 1/Dt, dBV, Ph
Automatic measurements: Simultaneously with waveform, 20 automatic measurements per channel and on the 4 channels simultaneously with scroll
MULTIMETER MODE
General specifications: 2 or 4 channels – 8,000 cts min/max/frequency/relative – TRMS – Time/date-stamped graphical recording in logger mode
AC, DC and AC + DC voltages: 600 mV to 600 VRMS, 800 mV to 800 VDC – VDC accuracy +/- (0.5 % + 25 D) – 200 kHz bandwidth
Resistance: 80 Ω to 32 MΩ – accuracy 0.5%R+ 25D – Quick continuity test < 10 ms
Other measurements: Temperature (HX0035 = KTC, HX0036 = Pt100) / Capacitance 5nF to 5mF / Frequency 200 kHz / Diode test 3.3 V
Single and three-phase power: Active, Reactive and Apparent power values plus Power Factor simultaneously with the U & I measurements
HARMONIC ANALYSER MODE
Multi-channel analysis: 2 or 4 (depending on model), 63 orders, fundamental frequency 40 to 450 Hz in auto or manual mode
Simultaneous measurements: Total Vrms, THD and selected order (% fundamental, phase, frequency, Vrms)
LOGGER MODE
Acquisition: Duration: 20,000 s – Interval: 0.2 s – Files: 100,000 measurements
GENERAL SPECIFICATIONS
Configuration memories: Not limited according to device - variable file sizes
Printing: Network printing via Ethernet/Wifi in .png format
PC communication – software link: Ethernet (100 baseT), WiFi-USB (device, 12 Mbs) – "ScopeNet" application software for PC
Software: PC: Ethernet and USB, ScopeNet (remote control, data recovery, cursors and automatic measurements)Android tablet – ScopeAdmin Fleet Administration utility
Mains power supply: Li-Ion rechargeable battery (6,900mAH-40 Wh) – Battery life of up to 8 hrs – Adjustable standby modeAdapter / 2-hour fast charger, universal 98-264 V / 50/60 Hz)
Safety / EMC: Safety as per IEC 61010-2-30, 2010 – 600V CAT III / 1000V CAT II – EMC as per EN61326-1, 2010
Mechanical specifications: 292.5 x 210.6 x 66.2 mm – 2.1 kg with batteries – IP54 protection
User Manual
Brochure
Technical Bulletin (p.13)